P102 HPLC pump, SS
PN: P102-011Description
It is recommended that the Self Flush feature be used to improve seal life in a number of applications. In particular, (as stated above) if pumping Buffers, Acids/Bases or any inorganic solution near saturation, the pump should utilize the Self Flush feature. With every piston stroke, an extremely thin film of solution is pulled back past the seal. If this zone is dry (without use of Self Flush) then crystals will form with continuous operation, which will ultimately damage the seal. Water or water/IPA mix is good choices for the flush solution.
|
Technická specifikace:
Flow Rate | 0,01 ml/min - 9,99 ml/min with increment 0,01 ml/min |
Max. pressure | 42 MPa (6.000 psi) |
Pressure limits | low and up - free settable |
Flow precision | 0,2%RSD (water @ 1 ml/min, p = min. 5 MPa) |
Flow accuracy | ± 2% |
Pulsation | ± 0,5% (water @ 1 ml/min, p = min. 5 MPa) |
Wetted materials | stainless steel 316, ceramic mat., inert polymer |
External communication | RS-232 |
Height x with x depth | 250 x 160 x 370 mm |
Weight | 9 kg |